Book chapter
Assessment of Surface Roughness and Waviness Using White Light Interferometry for Short-Term Implanted, Highly Crosslinked Acetabular Components
Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene for Joint Replacements, pp 41-56
01 Jan 2004
Featured in Collection : UN Sustainable Development Goals @ Drexel
Abstract
The reduction of volumetric wear continues to be one of the imminent challenges for the orthopaedic research community in the area of ultra-high molecular weight polyethylene (UHMWPE) joint replacement components. Retrieval analyses are necessary to determine the relationship between in vitro hip simulator predictions and actual wear performance in vivo. To quantify short term wear in retrieved highly crosslinked acetabular components it is necessary to differentiate between the initial surface morphology (dominated by manufacturer's machining marks) and the smaller scale surface features associated with in vivo adhesive/abrasive wear mechanisms. We have developed and validated a technique for deconvolution of as-manufactured versus in vivo generated surface topology from retrieved, highly crosslinked polyethylene acetabular inserts. Surface topology was characterized by white light interferometry with advanced texture analysis software. A Fourier transform algorithm was used to deconvolve the low-frequency features (i.e., waviness) such as machining marks, from the high-frequency features (i.e., roughness). Twenty-one (21) short-term (less than 24 months) conventional and highly crosslinked retrievals from different manufacturers were evaluated in this study. The wear surfaces in the short-term retrievals were deconvolved using the cut-off frequencies from the new inserts. The frequency distribution and magnitude of the machining marks were found to be material and manufacturer specific. This study highlights the importance of quantitative techniques, such as white light interferometry for distinguishing between initial and in vivo generated surface morphology. The topology observed in the crosslinked retrievals was consistent with the surface damage mechanisms previously observed in conventional UHMWPE components, namely macroscopic and microscopic adhesive/abrasive wear.
Metrics
Details
- Title
- Assessment of Surface Roughness and Waviness Using White Light Interferometry for Short-Term Implanted, Highly Crosslinked Acetabular Components
- Creators
- S M Kurtz - Exponent (United States)J Turner - Drexel UniversityM Herr - ExponentA A Edidin - Drexel UniversityC M Rimnac - Case Western Reserve University
- Publication Details
- Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene for Joint Replacements, pp 41-56
- Publisher
- ASTM International; 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959
- Number of pages
- 16
- Resource Type
- Book chapter
- Language
- English
- Academic Unit
- School of Biomedical Engineering, Science, and Health Systems; Infectious Diseases (and HIV Medicine)
- Web of Science ID
- WOS:000189423700004
- Scopus ID
- 2-s2.0-1342332176
- Other Identifier
- 991019173876404721
UN Sustainable Development Goals (SDGs)
This publication has contributed to the advancement of the following goals:
InCites Highlights
Data related to this publication, from InCites Benchmarking & Analytics tool:
- Web of Science research areas
- Engineering, Biomedical
- Materials Science, Biomaterials
- Orthopedics
- Polymer Science