Conference proceeding
Simulation and optimization of intra-aortic balloon pumping
[1989] Proceedings. Computers in Cardiology, pp 237-240
1989
Abstract
The interaction between the intra-aortic balloon pump and the cardiovascular system was studied to formulate a performance index and to design an optimal control system. An optimal control algorithm was selected and adapted to operate in a constrained time-varying environment. The optimization algorithm was implemented in a microprocessor-based controller which was successfully tested in animal experiments. Model predictions were also used to design a real-time interactive computer simulation to train intra-aortic balloon pump operators.< >
Metrics
16 Record Views
Details
- Title
- Simulation and optimization of intra-aortic balloon pumping
- Creators
- O Barnea - Tel Aviv UniversityB SmithT.W MooreD Jaron
- Publication Details
- [1989] Proceedings. Computers in Cardiology, pp 237-240
- Publisher
- IEEE Comput. Soc. Press
- Resource Type
- Conference proceeding
- Language
- English
- Academic Unit
- Information Science; School of Biomedical Engineering, Science, and Health Systems; [Retired Faculty]
- Other Identifier
- 991019182669104721