Logo image
Clinical, radiological, histological and retrieval findings of Activ-L and Mobidisc total disc replacements: a study of two patients
Journal article   Open access   Peer reviewed

Clinical, radiological, histological and retrieval findings of Activ-L and Mobidisc total disc replacements: a study of two patients

Shennah Austen, Ilona M. Punt, Jack P. M. Cleutjens, Paul C. Willems, Steven M. Kurtz, Daniel W. MacDonald, Lodewijk W. van Rhijn and André van Ooij
European spine journal, v 21(Suppl 4), pp 513-520
2012
PMID: 22245852
url
https://doi.org/10.1007/s00586-011-2141-7View
Published, Version of Record (VoR)Open Access (License Unspecified) Open

Abstract

Case Report Medicine Medicine & Public Health Neurosurgery Surgical Orthopedics
Introduction This study evaluates the short-term clinical outcome, radiological, histological and device retrieval findings of two patients with second generation lumbar total disc replacement (TDR). Materials and methods The first patient had a single level L4-L5 Activ-L TDR, the second patient a L4-L5 Mobidisc and L5-S1 Activ-L TDR. The TDRs were implanted elsewhere and had implantation times between 1.3 and 2.8 years. Results Plain radiographs and CT-scanning showed slight subsidence of the Activ-L TDR in both patients and facet joint degeneration. The patients underwent revision surgery because of recurrent back and leg pain. After removal of the TDR and posterolateral fusion, the pain improved. Histological examination revealed large ultrahigh molecular weight polyethylene (UHMWPE) particles and giant cells in the retrieved tissue surrounding the Mobidisc. The particles in the tissue samples of the Activ-L TDR were smaller and contained in macrophages. Retrieval analysis of the UHMWPE cores revealed evidence of minor adhesive and abrasive wear with signs of impingement in both TDR designs. Conclusion Although wear was unrelated to the reason for revision, this study demonstrates the presence of UHMWPE particles and inflammatory cells in second generation TDR. Long-term follow-up after TDR is indicated for monitoring wear and implant status.

Metrics

12 Record Views
25 citations in Scopus

Details

Logo image