Error Analysis for Vector Scope Measurement of Pulmonary Mechanics
Journal article   Peer reviewed

Error Analysis for Vector Scope Measurement of Pulmonary Mechanics

Peter A Koen, Gordon D Moskowitz and Thomas H Shaffer
IEEE transactions on biomedical engineering, v BME-25(2)
Mar 1978
PMID: 646865

Abstract

Biomedical measurements Calibration Electrical resistance measurement Equations Error analysis Immune system Lungs Measurement errors Mechanical variables measurement Oscilloscopes
The errors that are incurred by misreading the angle on the oscilloscope face in the determination of pulmonary compliance, resistance, and functional residual capacity with a vector scope are presented. Results indicate that the calculated value will be within 10% of the correct value, provided that the oscilloscope angle is between 10.5 and 78.5 degrees, and the reading error is less than 1 degree.