Feasibility of extravascular pacing with a novel substernal electrode configuration: The Substernal Pacing Acute Clinical Evaluation study
Journal article   Open access   Peer reviewed

Feasibility of extravascular pacing with a novel substernal electrode configuration: The Substernal Pacing Acute Clinical Evaluation study

Darius P. Sholevar, Stanley Tung, Vikas Kuriachan, Peter Leong-Sit, Henri Roukoz, Gregory Engel, Steven P. Kutalek, Devender Akula, Amy E. Thompson, Melissa G.T. Christie, …
Heart rhythm, v 15(4), pp 536-542
Apr 2018
PMID: 29197657
url
http://www.heartrhythmjournal.com/article/S1547527117314170/pdfView
Published, Version of Record (VoR) Open
url
https://doi.org/10.1016/j.hrthm.2017.11.030View
Published, Version of Record (VoR) Open

Abstract

Anterior mediastinum Clinical study Extravascular Pacing Substernal
Subcutaneous implantable cardioverter-defibrillators provide an alternative to transvenous defibrillation but require higher shock outputs and offer no antitachycardia pacing. The Substernal Pacing Acute Clinical Evaluation (SPACE) study evaluated the feasibility of pacing from an extravascular substernal location. The primary purpose of the SPACE study was to characterize pacing from the substernal space. Secondary objectives included evaluating extracardiac stimulation and recording electrograms. The SPACE study prospectively evaluated the feasibility of pacing with a commercially available electrophysiology catheter acutely implanted in the substernal space via minimally invasive subxiphoid access. Pacing data were collected in ≥7 vectors using constant current stimulation up to 20 mA and pulse width up to 10 ms. Catheter placement was successful in all 26 patients who underwent the procedure, with a mean placement time of 11.7 ± 10.1 minutes. Eighteen patients (69%) had successful ventricular capture in ≥1 tested vector. The mean pacing threshold at a pulse width of 10 ms was 7.3 ± 4.2 mA across all vectors (5.8 ± 4.4 V). Failed capture was generally associated with suboptimal catheter placement or presumed air ingression. A low level of extracardiac stimulation was observed in 1 patient. The mean R-wave amplitude ranged from 2.98 to 4.11 mV in the unipolar configuration and from 0.83 to 3.95 mV in the bipolar configuration. The data from the SPACE study demonstrate that pacing is feasible from the extravascular substernal location. A substernal electrode configuration has the potential to provide pacing in a future extravascular device without need for intracardiac hardware placement. [Display omitted]

Metrics

6 Record Views
35 citations in Scopus

Details

UN Sustainable Development Goals (SDGs)

This publication has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

InCites Highlights

Data related to this publication, from InCites Benchmarking & Analytics tool:

Collaboration types
Domestic collaboration
International collaboration
Web of Science research areas
Cardiac & Cardiovascular Systems
Logo image