Fretting Corrosion, Third-Body Polyethylene Damage, and Cup Positioning in Primary vs Revision Dual Mobility Total Hip Arthroplasty
The Journal of arthroplasty, v 36(7), pp S80-S87
Jul 2021
PMID: 33610405
Featured in Collection : UN Sustainable Development Goals @ Drexel
Abstract
Dual mobility (DM) articulations were introduced for total hip arthroplasty to reduce the risk of instability for patients who have a high risk of dislocation. The use of DM constructs in both primary and revision total hip arthroplasty has been steadily increasing, leading to concerns regarding potential risks of fretting corrosion, polyethylene wear, metal release, and failure due to component positioning.
A total of 56 retrieved DM constructs were collected. The inner and outer polyethylene liner surfaces were assessed for 7 damage mechanisms, and fretting corrosion was evaluated for the femoral stem, head, and modular liner. Three polyethylene liners with the greatest amounts of embedded debris were examined using scanning electron microscopy. Energy-dispersive X-ray spectroscopy was used to determine the elemental content of the debris. Acetabular cup orientation was analyzed radiographically using the EBRA (Einzel-Bild-Roentgen-Analyse) method.
The devices were revised most frequently for infection (36%), loosening (21%), and instability/dislocation (18%). The most common polyethylene damage mechanisms were scratching, pitting, burnishing, and embedded debris, and no difference in total damage was found between primary and revision cases. Scanning electron microscopy/energy-dispersive X-ray spectroscopy revealed that debris morphology and composition were consistent with porous titanium coating, resulting from cup loosening or broken screws and augments. A total of 71% and 50% of the constructs were determined to be within the Lewinnek safe zone for inclination and anteversion, respectively.
The most notable mechanisms of surface damage were due to third-body debris, especially for the polyethylene surfaces which articulate against cobalt-chromium femoral heads and acetabular liners. Scratching of the femoral head and the metal liner from this debris may support the clinical use of ceramic for DM bearing surfaces in the future.
Metrics
Details
- Title
- Fretting Corrosion, Third-Body Polyethylene Damage, and Cup Positioning in Primary vs Revision Dual Mobility Total Hip Arthroplasty
- Creators
- Hannah Spece - Drexel UniversityEric S. Ouellette - ExponentOlivia L. Jones - Drexel UniversityDaniel W. MacDonald - Drexel UniversityNicolas S. Piuzzi - Department of Orthopedic Surgery Cleveland Clinic Cleveland OHGwo-Chin Lee - Penn Presbyterian Medical Center, University of Pennsylvania School of Medicine, Department of Orthopaedic Surgery, Philadelphia, PAMichael A. Mont - Lenox Hill HospitalGregg R. Klein - Hackensack University Medical CenterSteven M. Kurtz - Exponent (United States)
- Publication Details
- The Journal of arthroplasty, v 36(7), pp S80-S87
- Publisher
- Elsevier
- Resource Type
- Journal article
- Language
- English
- Academic Unit
- School of Biomedical Engineering, Science, and Health Systems
- Web of Science ID
- WOS:000661072200062
- Scopus ID
- 2-s2.0-85101077063
- Other Identifier
- 991019168642904721
UN Sustainable Development Goals (SDGs)
This publication has contributed to the advancement of the following goals:
InCites Highlights
Data related to this publication, from InCites Benchmarking & Analytics tool:
- Collaboration types
- Industry collaboration
- Domestic collaboration
- Web of Science research areas
- Orthopedics