Fretting and Corrosion Damage in Taper Adapter Sleeves for Ceramic Heads: A Retrieval Study
Journal article   Peer reviewed

Fretting and Corrosion Damage in Taper Adapter Sleeves for Ceramic Heads: A Retrieval Study

Daniel W. MacDonald, Antonia F. Chen, Gwo-Chin Lee, Gregg R. Klein, Michael A. Mont, Steven M. Kurtz, Harold E. Cates, Matthew J. Kraay, Clare M. Rimnac and Taper Corrosion Writing Committee
The Journal of arthroplasty, v 32(9), pp 2887-2891
Sep 2017
PMID: 28668212

Abstract

femoral stem fretting corrosion revision hip arthroplasty sleeved ceramic head taper corrosion
During revision surgery with a well-fixed stem, a titanium sleeve can be used in conjunction with a ceramic head to achieve better stress distribution across the taper surface. In vitro testing suggests that corrosion is not a concern in sleeved ceramic heads; however, little is known about the in vivo fretting corrosion of the sleeves. The purpose of this study was to investigate fretting corrosion in sleeved ceramic heads in retrieved total hip arthroplasties. Thirty-seven sleeved ceramic heads were collected during revision. The femoral heads and sleeves were implanted 0.0-3.3 years. The implants were revised predominantly for instability, infection, and loosening. Fifty percent of the retrievals were implanted during a primary surgery. Fretting corrosion was assessed using the Goldberg-Higgs semiquantitative scoring system. Mild-to-moderate fretting corrosion scores (score = 2-3) were observed in 92% of internal tapers, 19% of external tapers, and 78% of the stems. Severe fretting corrosion was observed in 1 stem trunnion that was previously retained during revision surgery and none of the retrieved sleeves. There was no difference in corrosion damage of sleeves used in primary or revision surgery. The fretting corrosion scores in this study were predominantly mild and lower than reported fretting scores of cobalt-chrome heads in metal-on-polyethylene bearings. Although intended for use in revisions, we found that the short-term in vivo corrosion behavior of the sleeves was similar in both primary and revision surgery applications. From an in vivo corrosion perspective, sleeves are a reasonable solution for restoring the stem taper during revision surgery.

Metrics

4 Record Views
21 citations in Scopus

Details

UN Sustainable Development Goals (SDGs)

This publication has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

InCites Highlights

Data related to this publication, from InCites Benchmarking & Analytics tool:

Collaboration types
Industry collaboration
Domestic collaboration
Web of Science research areas
Orthopedics
Logo image