Improvement in plan quality after Implementation of clinical goals in a large network of cancer centers
Medical dosimetry : official journal of the American Association of Medical Dosimetrists, v 48(1), pp 51-54
01 Jan 2023
PMID: 36411200
Featured in Collection : UN Sustainable Development Goals @ Drexel
Abstract
Clinical Goals (CG) is a tool available in the Varian Eclipse planning system to objectively and visually evaluate the quality of treatment plans based upon user-defined dose-volume parameters. We defined a set of CG for Stereotactic Radiosurgery (SRS) and Intensity-Modulated Radiotherapy (IMRT) based on published data and guidelines and implemented this in a network of cancer centers in India (American Institute of Oncology). A dosimetric study was performed to compare brain SRS and breast IMRT plan quality before and after CG implementation.The CG defined for SRS plans were target V100% ≥ 98%, dose gradient measure (GM) ≤ 0.5 cm, conformity index (CI) 1.0 to 1.2. For breast IMRT plans, CG defined target V100% ≥ 97%, V95% ≥ 95%, V107% ≤ 2%, V105% ≤ 10%, and Dmax ≤ 2.4 Gy. Dose limits to organs-at-risk (OAR) were summarize in supplemental materials. Twenty brain SRS and 10 breast IMRT treatment plans that were previously delivered on patients were selected and re-planned using CG. The pre and postoptimized plan parameters were compared using student t-tests.For brain SRS plans, the V100, GM, and CI for the pre- and post-Clinical-Goals plans were 93.22% ± 7.2% vs 97.96% ± 0.29% (p = 0.009), 0.63 ± 0.16 vs 0.42 ± 0.05 (p < 0.001) and 1.07 ± 0.18 vs 1.06 ± 0.06 (p = 0.79), respectively. There were no differences in max dose to OARs. In breast IMRT plans, the target V107% for pre and postimplemented plans were 16.50% ± 10.98% vs 0.32% ± 0.32%, respectively (p = 0.001). The average target V105% were 44.00% ± 15.72% and 8.69% ± 4.53%, respectively (p < 0.001). No differences were found in the average target V100% (p = 0.128) and V95% (p = 0.205). The average target Dmax were 112.28% ± 1.59% and 109.14% ± 0.73%, respectively (p < 0.001). There were only minor differences in doses to OARs.The implementation of CG in Varian Eclipse significantly improved SRS and IMRT plan quality with enhanced coverage, dose GM, and CI without increased dose to OARs.
Metrics
Details
- Title
- Improvement in plan quality after Implementation of clinical goals in a large network of cancer centers
- Creators
- Hefei Liu - Varian Medical SystemsPraveen Nuksani - Varian Medical SystemsMalolan Rajagopalan - Mount Carmel HealthMangesh Patil - Basavatarakam Indo American Cancer Hospital and Research InstituteKrishna Komanduri - Varian Medical SystemsBrent Murphy - Varian Medical SystemsDeepak Khuntia - Varian Medical SystemsSushil Beriwal - Varian Medical Systems
- Publication Details
- Medical dosimetry : official journal of the American Association of Medical Dosimetrists, v 48(1), pp 51-54
- Publisher
- Elsevier
- Resource Type
- Journal article
- Language
- English
- Academic Unit
- Radiation Oncology (and Nuclear Medicine)
- Web of Science ID
- WOS:000927793600001
- Scopus ID
- 2-s2.0-85142489059
- Other Identifier
- 991021897271004721
UN Sustainable Development Goals (SDGs)
This publication has contributed to the advancement of the following goals:
InCites Highlights
Data related to this publication, from InCites Benchmarking & Analytics tool:
- Collaboration types
- Domestic collaboration
- International collaboration
- Web of Science research areas
- Oncology
- Radiology, Nuclear Medicine & Medical Imaging