Interlaboratory acoustic power measurement

Peter A Lewin, Nadine Barrie-Smith, Masao Ide, Kullervo Hynynen and Michael Macdonald
Journal of ultrasound in medicine, v 22(2), pp 207-213
Feb 2003
PMID: 12562125

Abstract

Biological and medical sciences Investigative techniques, diagnostic techniques (general aspects) Medical sciences Miscellaneous. Technology Ultrasonic investigative techniques
Objective. This article describes an American Institute of Ultrasound in Medicine–sponsored intercomparison of the results of acoustic power measurements performed by several laboratories. Methods. Two primary calibration techniques, namely, planar scanning and radiation force balance, were used in the frequency range typical of that in which sonographic imaging devices operate. The same reference source, the National Institute of Standards and Technology (Gaithersburg, MD) standard ultrasonic power source, capable of producing acoustic fields in the frequency range from approximately 1 to 21 MHz, was circulated to 3 laboratories. Results. The results of the calibrations indicate that the overall uncertainty in acoustic power measurements depends on the target and the measurement method. In the case of radiation force balance measurements with an absorbing target, the largest discrepancy between the available National Institute of Standards and Technology–calibrated results and the reported data was 10.6% at approximately 2.5 MHz. At higher frequencies, beyond 10 MHz, the largest discrepancy reported with an absorbing target was 8.4%. For a reflecting target, the largest discrepancies were 16.2% at approximately 3.7 MHz and 15.4% at about 10 MHz. The largest discrepancy identified for the planar scanning technique below 10 MHz was 7.4% at 3.7 MHz. Conclusions. The results obtained suggest that an absorbing target may be preferable for acoustic power measurements with radiation force balance. In a group that consists of 2 research laboratories and 1 manufacturer, the power measurements agreed within 16%.

Metrics

8 Record Views
17 citations in Scopus

Details

UN Sustainable Development Goals (SDGs)

This publication has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

InCites Highlights

Data related to this publication, from InCites Benchmarking & Analytics tool:

Collaboration types
Industry collaboration
Domestic collaboration
International collaboration
Web of Science research areas
Acoustics
Radiology, Nuclear Medicine & Medical Imaging
Logo image