Journal article
Load Sharing and Stabilization Effects of Anterior Cervical Devices
Journal of spinal disorders & techniques, v 22(8), pp 571-577
01 Dec 2009
PMID: 19956031
Featured in Collection : UN Sustainable Development Goals @ Drexel
Abstract
Study Design: Biomechanical human cadaveric cervical flexibility testing with direct load-sharing measurement.
Objective: To determine if the flexibility and load-sharing characteristics of a functional spinal unit were affected by anterior treatments for cervical pathologies.
Summary of Background Data: With advancements in polymers, anterior cervical plates have used thermoplastics including recent designs from biodegradable polylactide acids. However, the difference in material properties between metal and polymer can be significant.
Methods: Thirteen cervical spine specimens were subjected to 5 treatments at C4-C5. Each treatment for each specimen was subjected to multidirectional flexibility testing. The third cycle was used for treatment comparisons.
Results: With the integrated load cell spacer, the mean range of motion for the functional spinal unit measured on average 104% +/- 40% normalized to the intact control. The mean biodegradable and titanium plate were 55% +/- 31% and 40% +/- 36%, respectively. Both plates exhibited statistically lower mean range of motions (P = 0.001 and P < 0.001) compared with spacers. The load transmitted through the interbody space was 54% +/- 20%, 43% +/- 20%, and 33% +/- 15% on average for the spacer, biodegradable, and titanium plate constructs, respectively. No statistically significant difference was detected between the biodegradable plate and spacer (P = 0.214).
Conclusions: From this research, a biodegradable plate offers immediate postoperative stability significantly different than spacer alone treatments but with graft load sharing that is statistically no different. Thus, the intrinsic lower native material modulus of elasticity of biodegradable polymers has biomechanical implications. However, clinical evidence, particularly for long-term outcomes, will be required in understanding the efficacy of biodegradable polymers.
Metrics
Details
- Title
- Load Sharing and Stabilization Effects of Anterior Cervical Devices
- Creators
- Boyle C. Cheng - Neurological SurgeryPaul Burns - Synthes Spine, Philadelphia, PA USAStephen Pirris - Neurological SurgeryWilliam C. Welch - University of Pennsylvania
- Publication Details
- Journal of spinal disorders & techniques, v 22(8), pp 571-577
- Publisher
- Lippincott Williams & Wilkins
- Number of pages
- 7
- Resource Type
- Journal article
- Language
- English
- Academic Unit
- Surgery
- Web of Science ID
- WOS:000279665500006
- Scopus ID
- 2-s2.0-74249109944
- Other Identifier
- 991021930321604721
UN Sustainable Development Goals (SDGs)
This publication has contributed to the advancement of the following goals:
InCites Highlights
Data related to this publication, from InCites Benchmarking & Analytics tool:
- Collaboration types
- Domestic collaboration
- Web of Science research areas
- Clinical Neurology
- Orthopedics