Background:
The efficacy of using diffuse near infrared spectroscopy (NIRS) in predicting wound healing in diabetic foot ulcers was demonstrated by conducting a pilot human study.
Method:
Sixteen chronic diabetic wounds were followed and assessed for subsurface oxyhemoglobin concentration using the NIRS device. Weekly measurements were conducted until there was wound closure, limb amputation, or 20 completed visits without healing. Wound size and degree of wound contraction were measured by image analysis of digital photographs, and results were compared to NIRS results.
Results:
In the 16 patients followed, seven wounds healed, six limbs were amputated, and three wounds remained opened after 20 visits. Initial values of subsurface hemoglobin concentration, in all wounds, were higher than in nonwound control sites. Healed wounds exhibited a consistent reduction of hemoglobin concentration several weeks prior to closure, and the absolute hemoglobin concentration approached the value at the control site. In wounds that did not heal or ended in amputations, the hemoglobin concentration remained elevated throughout the study. A negative slope for the rate of change of hemoglobin concentration was indicative of healing across all wounds.
Conclusions:
Evaluation of diabetic foot ulcers using NIRS may provide an effective and more complete measurement of wound healing compared to current clinical approaches.
Metrics
7 Record Views
40 citations in Scopus
Details
Title
Near Infrared Wound Monitor Helps Clinical Assessment of Diabetic Foot Ulcers
Creators
Michael Neidrauer - Drexel University
Leonid Zubkov - Drexel University
Michael S. Weingarten - Drexel University
Kambiz Pourrezaei - Drexel University
Elisabeth S. Papazoglou - Drexel University
Publication Details
Journal of diabetes science and technology, v 4(4), pp 792-798
Publisher
Sage
Resource Type
Journal article
Language
English
Academic Unit