Abstract

Diagnostic Techniques, Neurological - instrumentation Equipment Failure Female Humans Male Temperature Ventriculoperitoneal Shunt
While a noninvasive flow determination would be desirable in the diagnosis of cerebrospinal fluid shunt malfunction, existing studies have not yet defined a role for thermal flow detection. To evaluate a revised test protocol using a micropumper designed to transiently enhance flow during thermal testing to determine whether thermal detection of flow is associated with progression to shunt revision surgery. Eighty-two unique tests were performed in 71 shunts. The primary outcome, need for revision within 7 d of testing, was compared with results of micropumper-augmented thermal flow detection. Statistical analysis was based on blind interpretation of test results and raw temperature data recorded during testing. The test was sensitive (73%) and specific (68%) in predicting need for revision, with 5.6-fold higher probability of revision when flow was not detected. Negative predictive value in our sample was 94.2%. The probability of not requiring revision increased with increasing total temperature drop. Analysis of various possible thresholds showed that the optimal temperature cutoff may be lower than suggested by the manufacturer (0.125°C vs 0.2°C). This is the first study to report a strong association between thermal flow evaluation and a clinical impression that a shunt is not malfunctioning. The current recommended threshold may increase the false positive rate unnecessarily, and as clinicians gain experience with the method, they may find value in examining the temperature curves themselves. Multicenter studies are suggested to further define a role for this diagnostic test.

Metrics

19 Record Views
8 citations in Scopus

Details

UN Sustainable Development Goals (SDGs)

This publication has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

InCites Highlights

Data related to this publication, from InCites Benchmarking & Analytics tool:

Collaboration types
Domestic collaboration
Web of Science research areas
Clinical Neurology
Surgery
Logo image