Oxidation, Damage Mechanisms, and Reasons for Revision of Sequentially Annealed Highly Crosslinked Polyethylene in Total Knee Arthroplasty
Journal article   Open access   Peer reviewed

Oxidation, Damage Mechanisms, and Reasons for Revision of Sequentially Annealed Highly Crosslinked Polyethylene in Total Knee Arthroplasty

Daniel W. MacDonald, Genymphas B. Higgs, Antonia F. Chen, Arthur L. Malkani, Michael A. Mont and Steven M. Kurtz
The Journal of arthroplasty, v 33(4), pp 1235-1241
Apr 2018
PMID: 29329883
url
https://doi.org/10.1016/j.arth.2017.09.036View
Published, Version of Record (VoR)CC BY-NC-ND V4.0 Open

Abstract

highly crosslinked polyethylene HXLPE oxidation reasons for revision total knee arthroplasty
Sequentially annealed, highly crosslinked polyethylene (HXLPE) has been used clinically in total knee arthroplasty (TKA) for over a decade. However, little is known about the revision reasons; its surface damage mechanisms; or its in vivo oxidative stability relative to conventional polyethylene. We asked whether retrieved HLXPE tibial inserts exhibited: (1) similar revision reasons; (2) improved resistance to surface damage; and (3) improved oxidative stability, when compared with conventional gamma inert sterilized polyethylene inserts. A total of 456 revised tibial inserts were collected in a multicenter retrieval program between 2000 and 2016. The implantation time for the HXLPE components was 1.8 ± 1.8 years, and for the control inserts it was 3.4 ± 2.7 years. Revision reasons were assessed based on medical records, radiographs, and examinations of the retrieved components. Surface damage was assessed using a semi-quantitative scoring method. Oxidation was measured using Fourier transform infrared spectroscopy. The tibial inserts in both cohorts were revised most frequently for loosening, infection, and instability. The most commonly observed surface damage modes were burnishing, pitting, and scratching. Oxidation of the HXLPE inserts was, on average, low and similar to the control inserts at the bearing surface and the stabilizing post. We observed evidence of in vivo oxidation in both HXLPE and control tibial inserts. We found no association between the levels of oxidation and the clinical performance of the HXLPE tibial components. The findings of this study document the revision reasons, surface damage modes, and oxidative behavior of sequentially annealed HXLPE for TKA.

Metrics

5 Record Views
22 citations in Scopus

Details

UN Sustainable Development Goals (SDGs)

This publication has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

InCites Highlights

Data related to this publication, from InCites Benchmarking & Analytics tool:

Collaboration types
Industry collaboration
Domestic collaboration
Web of Science research areas
Orthopedics
Logo image