Journal article
Percutaneous Extraction of ePTFE-Coated ICD Leads: A Single Center Comparative Experience
Pacing and clinical electrophysiology, v 36(4), pp 444-450
01 Apr 2013
PMID: 23330676
Featured in Collection : UN Sustainable Development Goals @ Drexel
Abstract
Background Percutaneous extraction of standard implantable cardioverter-defibrillator leads is often complicated by ingrowth of fibrotic tissue into the shocking coils. Leads with GORE expanded polytetrafluoroethylene (ePTFE) coating (W. L. Gore & Associates, Inc., Newark, DE, USA) designed to inhibit fibrosis are in use, but clinical data regarding their extraction are lacking. The study's purpose was to examine the feasibility, efficacy, and safety of percutaneous extraction involving defibrillator leads coated with ePTFE. Methods We analyzed our database to identify all percutaneously extracted leads with ePTFE-coated shocking coils. Lead and procedure characteristics were compared to a cohort of noncoated leads of similar implant duration. Results One hundred fifty-six leads were extracted from 145 patients; 57 ePTFE-coated leads, with a mean implant duration of 621 days, were extracted and compared to 99 noncoated leads, with a mean implant duration of 763 days (P = 0.0641). Mean extraction time was 5 minutes for coated leads versus 9.75 minutes for noncoated leads (P = 0.0001). Extraction time of less than 1 minute was more frequent with coated leads (61% vs 35%, P = 0.0025). Adjunct extraction tools were required less frequently with coated leads than noncoated leads (39% vs 63%, P = 0.0071). There was no fibrosis where ePTFE covered the shocking coils. Alternatively, 23 of 99 (23%) noncoated leads demonstrated fibrosis adherent to the shock coil. There were no procedure-related complications in either group. Conclusions Compared to noncoated leads, ePTFE-coated leads are associated with shorter extraction times and are less likely to require extraction tools for removal. The difference is likely related to the absence of fibrosis over the ePTFE-coated high-energy coils. (PACE 2013; 36:444-450)
Metrics
Details
- Title
- Percutaneous Extraction of ePTFE-Coated ICD Leads: A Single Center Comparative Experience
- Creators
- Andrew R. Kohut - Drexel Univ, Div Cardiol, Dept Med, Coll Med, Philadelphia, PA 19102 USAJon Grammes - Drexel Univ, Div Cardiol, Dept Med, Coll Med, Philadelphia, PA 19102 USAChristopher M. Schulze - Cardiol Consultants Philadelphia, Philadelphia, PA USAMohammad Al-Bataineh - Drexel UniversityGeorge A. Yesenosky - Cardiology Consultants of Philadelphia Philadelphia PennsylvaniaJay C. Horrow - Drexel UniversitySteven P. Kutalek - Drexel University
- Publication Details
- Pacing and clinical electrophysiology, v 36(4), pp 444-450
- Publisher
- Wiley
- Number of pages
- 7
- Resource Type
- Journal article
- Language
- English
- Academic Unit
- College of Medicine; Anesthesiology and Perioperative Medicine
- Web of Science ID
- WOS:000317020600014
- Scopus ID
- 2-s2.0-84875805824
- Other Identifier
- 991019169539404721
UN Sustainable Development Goals (SDGs)
This publication has contributed to the advancement of the following goals:
InCites Highlights
Data related to this publication, from InCites Benchmarking & Analytics tool:
- Web of Science research areas
- Cardiac & Cardiovascular Systems
- Engineering, Biomedical