Polyethylene Stresses in Lumbar Total Joint Replacement Under Elevated Loading: Insights from an Anatomic Finite Element Model
Journal article   Open access   Peer reviewed

Polyethylene Stresses in Lumbar Total Joint Replacement Under Elevated Loading: Insights from an Anatomic Finite Element Model

Hannah Spece
Bioengineering, v 13(1), 66
06 Jan 2026
PMID: 41595998
Featured in Collection :   Research Supported by Drexel Libraries' OA Programs
url
https://doi.org/10.3390/bioengineering13010066View
Published, Version of Record (VoR)Open Access Discount via Drexel Libraries Read and Publish Program 2025CC BY V4.0 Open

Abstract

spine arthroplasty elevated loading finite element model highly crosslinked polyethylene wear contact pressure von Mises stress
The goal of this study was to assess elevated spinal loading conditions and their effect on the polyethylene stresses of a lumbar total joint replacement (L-TJR). A previously validated lumbar spine finite element model was virtually implanted with an L-TJR at L4–L5 and exposed to three elevated loading conditions: (1) 95th-percentile male body weight while bending forward, (2) combined ±7.5 Nm axial torsion and lateral bending, and (3) ASTM F2423 aggressive loading (1850 N plus 10–12 Nm bending). Combined torsion and lateral bending were considered because these loads and moments may be coupled in demanding real-world scenarios. Across all conditions, contact at the bearing remained confined to the intended spherical surfaces, consistent with Mode I in vitro wear tests, with no evidence of impingement. Contact stresses and von Mises stresses were considered acceptable based on the simulated results of Mode IV impingement tests. Only in one scenario—95th-percentile male body weight with multiaxial torsion—did von Mises stress in the polyethylene slightly exceed the stresses associated with impingement (<5%). These findings are useful in establishing the upper biomechanical loading limits for the L-TJR design beyond the 50th-percentile loading levels employed by standard in vitro tests. Future validation efforts such as a comparison with retrieval analyses or clinical data will further strengthen the model’s applicability to current and future questions of interest and contexts of use. Additional work may expand the modeling framework to incorporate patient-specific anatomy, variable implant positioning conditions, and a broader range of physiological load scenarios.

Metrics

3 Record Views

Details

UN Sustainable Development Goals (SDGs)

This publication has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

InCites Highlights

Data related to this publication, from InCites Benchmarking & Analytics tool:

Web of Science research areas
Biotechnology & Applied Microbiology
Engineering, Biomedical
Logo image