Pulse Oximetry in Sick Premature Infants and the Effects of Phototherapy and Radiant Warmers on the Oxygen Saturation Readout
American journal of perinatology, v 7(1)
Jan 1990
PMID: 2294914
Featured in Collection : UN Sustainable Development Goals @ Drexel
Abstract
ABSTRACT
Forty-five arterial blood samples for oxygen saturation on 15 sick premature infants were significantly correlated with pulse oximeter determinations (r = 0.87, p < 0.0001) giving a regression equation of y = 1.19X-18.15. Chronologic age in the first 2 weeks of life did not alter the accuracy of the pulse oximeter.
Four groups (in isolette with phototherapy off, in isolette with phototherapy on, on radiant warmer with phototherapy off, and on radiant warm with phototherapy on) were observed to see if environmental light and energy would affect pulse oximetry oxygen saturation values; no effect was observed.
Therefore the pulse oximeter is an accurate machine to assess sick premature infant oxygen saturation. The measurements are not altered by environmental light.
Metrics
Details
- Title
- Pulse Oximetry in Sick Premature Infants and the Effects of Phototherapy and Radiant Warmers on the Oxygen Saturation Readout
- Creators
- Alan B. Zubrow - Drexel UniversityGinny W. HendersonSonia O. ImaizumiJeanette R. Pleasure
- Publication Details
- American journal of perinatology, v 7(1)
- Publisher
- Thieme
- Resource Type
- Journal article
- Language
- English
- Academic Unit
- Pediatrics
- Web of Science ID
- WOS:A1990CH86700019
- Scopus ID
- 2-s2.0-0025157289
- Other Identifier
- 991019184299504721
UN Sustainable Development Goals (SDGs)
This publication has contributed to the advancement of the following goals:
InCites Highlights
Data related to this publication, from InCites Benchmarking & Analytics tool:
- Web of Science research areas
- Obstetrics & Gynecology
- Pediatrics