Journal article
Radiation exposure to hand surgeons' hands: a practical comparison of large and mini C-arm fluoroscopy
The Journal of hand surgery (American ed.), v 39(9), pp 1805-1809
01 Sep 2014
PMID: 25086796
Featured in Collection : UN Sustainable Development Goals @ Drexel
Abstract
To determine and compare the radiation exposure to surgeons' hands with large and mini C-arm fluoroscopy in a practical, clinically based model.
Two hand surgeons monitored radiation exposure to their hands with a ring dosimeter over a 14-month period using large and mini C-arm fluoroscopic units. One surgeon performed all cases with a large C-arm unit in a hospital setting, and the other performed all cases with mini C-arms in surgical centers. For each case, fluoroscopic time, the output displayed by the unit, radiation by time, and ring dosimeter absorption were recorded and analyzed.
A total of 160 consecutive cases were reviewed with 71 cases and 89 cases in the large and mini C-arm groups, respectively. The median output displayed by the large C-arm was 0.7 mGy/case, and the median output displayed by the mini C-arm was 10.0 mGy/case. With output as a product of time, the median calculated values were 0.02 mGy/s for the large C-arm group and 0.28 mGy/s for the mini C-arm group. Cumulative ring dosimeter absorption to the surgeons' hands was found to be 380 mrem for 71 cases in the large C-arm group versus 1,000 mrem for 89 cases in the mini C-arm group.
In our model, the use of the mini C-arm resulted in more than a 10-fold increase in the rate of output and approximately double the dosimeter absorption to the surgeon's hand compared with the large C-arm. Although it has been shown that the mini C-arm produces less radiation scatter, in a practical model, it may not be a safer alternative with respect to the surgeon's hands. Based on these findings, we recommend that surgeons be more aware of radiation exposure risk, know their C-arm unit's specifications, and try to minimize radiation exposure.
Prognostic II.
Metrics
Details
- Title
- Radiation exposure to hand surgeons' hands: a practical comparison of large and mini C-arm fluoroscopy
- Creators
- Michael M Vosbikian - Thomas Jefferson University HospitalAsif M Ilyas - Rothman InstituteDerek D Watson - Thomas Jefferson University HospitalCharles F Leinberry - Thomas Jefferson University Hospital
- Publication Details
- The Journal of hand surgery (American ed.), v 39(9), pp 1805-1809
- Publisher
- Elsevier
- Resource Type
- Journal article
- Language
- English
- Academic Unit
- Pediatrics
- Web of Science ID
- WOS:000342036500022
- Scopus ID
- 2-s2.0-84906937453
- Other Identifier
- 991021838155204721
UN Sustainable Development Goals (SDGs)
This publication has contributed to the advancement of the following goals:
InCites Highlights
Data related to this publication, from InCites Benchmarking & Analytics tool:
- Collaboration types
- Domestic collaboration
- Web of Science research areas
- Orthopedics
- Surgery