Abstract

Alumina Ceramic Femoral head Fracture Proof testing Review Strength Total hip arthroplasty
Ceramic components are used increasingly in total hip arthroplasty (THA). Compared with metallic components, ceramic femoral heads for THA have the potential advantage of lower wear rates in articulations with acetabular liners. However, the use of ceramic components is also associated with unique risks, including sudden fracture and intolerable bearing noise or squeaking. This review paper summarizes the published literature regarding alumina ceramic femoral heads and tries to identify areas where uncertainties remain. We will discuss the following topics: (1) the fracture mechanics of ceramic materials; (2) design-related stresses acting on the femoral head, especially at the interface between the stem trunnion and the head; (3) gradual loss of strength in service by fatigue or slow crack growth and simulation using a proof test; and (4) information that can be gathered from examination of fractured explants.

Metrics

10 Record Views
58 citations in Scopus

Details

UN Sustainable Development Goals (SDGs)

This publication has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

InCites Highlights

Data related to this publication, from InCites Benchmarking & Analytics tool:

Collaboration types
Industry collaboration
Domestic collaboration
Web of Science research areas
Engineering, Biomedical
Materials Science, Biomaterials
Logo image