Logo image
The KamLAND full-volume calibration system
Journal article   Open access   Peer reviewed

The KamLAND full-volume calibration system

B. E. Berger, J. Busenitz, T. Classen, M. P. Decowski, D. A. Dwyer, G. Elor, A. Frank, S. J. Freedman, B. K. Fujikawa, M. Galloway, …
Journal of instrumentation, v 4(4), pp P04017-P04017
01 Apr 2009
url
http://citeseerx.ist.psu.edu/viewdoc/summary?doi=10.1.1.244.1848View
url
https://doi.org/10.1088/1748-0221/4/04/P04017View
Published, Version of Record (VoR) Open

Abstract

Instruments & Instrumentation Science & Technology Technology
We have successfully built and operated a source deployment system for the Kam-LAND detector. This system was used to position radioactive sources throughout the delicate 1-kton liquid scintillator volume, while meeting stringent material cleanliness, material compatibility, and safety requirements. The calibration data obtained with this device were used to fully characterize detector position and energy reconstruction biases. As a result, the uncertainty in the size of the detector fiducial volume was reduced by a factor of two. Prior to calibration with this system, the fiducial volume was the largest source of systematic uncertainty in measuring the number of anti-neutrinos detected by KamLAND. This paper describes the design, operation and performance of this unique calibration system.