Steven M. Kurtz, Avram A. Edidin and Donald L. Bartel
Journal of biomechanics, v 30(6), pp 639-642
1997
PMID: 9165399
url
https://doi.org/10.1016/s0021-9290(96)00181-9View
Published, Version of Record (VoR)CC BY-NC-ND V4.0 Open
url
https://doi.org/10.1016/S0021-9290(96)00181-9View
Published, Version of Record (VoR) Open

Abstract

Acetabular component Contact stress Metal backing Polishing, Cup orientation
Mechanical interactions between the polyethylene liner and the metal-backing play an important role in the load transfer and debris-generation mechanisms of an acetabular component. Insert thickness, cup orientation, and insert-shell interface conditions affect the resulting contact stresses at the articulating and backside surfaces of the polyethylene component. The objective of this study was to determine the variation in contact stresses on a hemispherical acetabular component as a function of the friction coefficient of the liner-shell interface, the thickness of the insert, and the load application angle. Three-dimensional finite element models of a metal-backed acetabular component with liner thicknesses of 3–12 mm were developed. The insert-shell interface was modeled as either matte or highly polished, and the load angle of the joint reaction force was changed from 36 to 63° with respect to the dome. We found that the contact stresses at the articulating and backside surfaces of the insert were relatively insensitive to changes in the coefficient of friction at the insert-shell interface (resulting in ∼1–10% variation in contact stress), when compared to the effect of changing the insert's thickness (∼80% variation in contact stress) or changing the direction of the joint reaction force (∼20% variation in contact stress). The results of this study suggest that polishing the metal at the insert-shell interface does not substantially change the contact stresses at either surface of the component. Of the design variables available for selective modification by either the surgeon or the engineer, insert thickness and shell orientation play a greater role in determining the magnitude of the resulting contact stresses.

Metrics

10 Record Views
53 citations in Scopus

Details

UN Sustainable Development Goals (SDGs)

This publication has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

InCites Highlights

Data related to this publication, from InCites Benchmarking & Analytics tool:

Web of Science research areas
Biophysics
Engineering, Biomedical
Logo image