As transcranial ultrasound stimulation (TUS) advances as a precise,
non-invasive neuromodulatory method, there is a need for consistent reporting
standards to enable comparison and reproducibility across studies. To this end,
the International Transcranial Ultrasonic Stimulation Safety and Standards
Consortium (ITRUSST) formed a subcommittee of experts across several domains to
review and suggest standardised reporting parameters for low intensity TUS,
resulting in the guide presented here. The scope of the guide is limited to
reporting the ultrasound aspects of a study. The guide and supplementary
material provide a simple checklist covering the reporting of: (1) the
transducer and drive system, (2) the drive system settings, (3) the free field
acoustic parameters, (4) the pulse timing parameters, (5) \emph{in situ}
estimates of exposure parameters in the brain, and (6) intensity parameters.
Detailed explanations for each of the parameters, including discussions on
assumptions, measurements, and calculations, are also provided.
Metrics
10 Record Views
Details
Title
ITRUSST Consensus on Standardised Reporting for Transcranial Ultrasound Stimulation
Creators
Eleanor Martin
Jean-François Aubry
Mark Schafer - Drexel University
Lennart Verhagen
Bradley Treeby
Kim Butts Pauly
Publication Details
arXiv.org
Resource Type
Preprint
Language
English
Academic Unit
School of Biomedical Engineering, Science, and Health Systems
Other Identifier
991021854471204721
Research Home Page
Browse by research and academic units
Learn about the ETD submission process at Drexel
Learn about the Libraries’ research data management services