Twisted cardiovascular cages for intravascular axial flow blood pumps to support the Fontan physiology
Journal article   Peer reviewed

Twisted cardiovascular cages for intravascular axial flow blood pumps to support the Fontan physiology

Amy L Throckmorton, Emily A Downs, John A Hazelwood, Jonathan O Monroe and Steven G Chopski
International journal of artificial organs, v 35(5), pp 369-375
May 2012
PMID: 22661112

Abstract

Heart Defects, Congenital - surgery Models, Cardiovascular Humans Heart Failure - physiopathology Ventricular Function Stress, Mechanical Hydrodynamics Materials Testing Fontan Procedure - adverse effects Models, Statistical Prosthesis Design Heart Failure - therapy Regression Analysis Biomechanical Phenomena Diffusion Chambers, Culture Numerical Analysis, Computer-Assisted Computer Simulation Adolescent Blood Flow Velocity Adult Heart Defects, Congenital - physiopathology Ventricular Pressure Heart Failure - etiology Heart-Assist Devices
Failing single ventricle physiology represents an ongoing challenge in mechanical assist device development, requiring pressure augmentation in the cavopulmonary circuit, reduction of systemic venous pressure, and increased cardiac output to achieve hemodynamic stabilization. To meet these requirements, we are developing a percutaneously-placed, axial flow blood pump to support ailing single ventricle physiology in adolescents and adults. We have modified the outer cage of the device to serve as both a protective and functional design component. This study examined the performance of 3 cage geometries with varying directions of filament twist using numerical simulations and hydraulic experiments. All 3 cage and pump models performed in acceptable ranges to support Fontan patients. The cage design employing filaments that are twisted in the opposite direction to the impeller blades and in the direction of the diffuser blades (against-with) demonstrated superior performance by generating a pressure rise range of 5-38 mmHg of flow rates of 0.5-6 l/min at rotational speeds of 5000-7000 rpm. The blood damage indices for all of the cages were found to be well below 2%, and the scalar stress levels were below 200 Pa. This study represents ongoing progress in the development of the impeller and cage assembly. Validation of the results will continue in experiments with blood bag evaluation as well as by particle image velocimetry measurements.

Metrics

7 Record Views
8 citations in Scopus

Details

UN Sustainable Development Goals (SDGs)

This publication has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

InCites Highlights

Data related to this publication, from InCites Benchmarking & Analytics tool:

Web of Science research areas
Engineering, Biomedical
Transplantation
Logo image